Article

lock Open Access lock Peer-Reviewed

71

Views

ORIGINAL ARTICLE

Analysis of the early and late use of the intra-aortic balloon pump as a method of mechanical support for the management of patients with low output syndrome, following cardiac surgery with cardiopulmonary bypass

Hugo de Moraes Sarmento MacruzI; Oscar Howard FrazierII; Denton A CooleyII

DOI: 10.1590/S0102-76381995000400004

ABSTRACT

The purpose of this study was to evaluate the intra-aortic balloon pump as a method of mechanical support for the management of patients with low output syndrome, following cardiac surgery with cardiopulmonary bypass. The results with early (intra-operative) and late (post-operative) support were compared, as well as the relationship among cardiopulmonary bypass time, aortic clamp time and mortality. Pre-operative left ventricle ejection fraction, as a prognosis index, and complications rates were also analysed. One hundred and thirty patients from Texas Heart Institute, during the period of January to December 1987, were studied retrospectively; there were 103 men and 27 women, with a mean age of 61.5 ± 10.7 (14 to 84) years. The mean weight body and body surface area were 75.5 ± 16.6 (42 to 134) kilograms and 1.87 ± 0.24 (1.08 to 2.60) square meters respectively. The mean time of cardiopulmonary bypass was 98 (33 to 299) minutes and the mean time of aortic clamp was 49 (10 to 122) minutes. All the patients had low cardiac output, associated with refractary arrythmias in 4 and after cardiac arrest in the intensive care unit in 2 patients. The intra-aortic balloon pump was left for variable periods of time from 15 minutes to 256 hours. The all cohort was divided in six groups, according to the early (intra-operative) insertion of the balloon pump (Group I), the late (post-operative) insertion (Group II), patient's age equal or inferior to 65 years Group III and Group IV age superior to 65 years, Group V and VI according cardiopulmonary bypass time up to 120 or superior to 120 minutes, and both were divided according to aortic clamp time (subgroup 1, up to 60 minutes and subgroup 2, superior to 60 minutes). The outcome was successful in 81 (62.3%) and 49 (37.7%) died (36.2% from Group I and 44% from Group II). The survival rate was greater in Group III (68.6%) than in Group IV (51 %) with p<0.05 showing statistical relevance of mortality in the older group. There were no significant associations between ejection fraction, aortic clamp time versus mortality but cardiopulmonary bypass had high statistical significance with p<0.01 for times over 120 minutes. Complications were found in 6 patients (4.6%). These data suggest that intra-aortic balloon pump is an effective mechanical support for the management of patients with low cardiac output following cardiac surgery with cardiopulmonary bypass, showing a tendency of best results when earliest applied, there is an influence of age and time of cardiopulmonary bypass over the mortality rates, suggesting that in these patients should be used more aggressive and effective circulatory support device.

RESUMO

Este estudo tem como objetivo, a análise retrospectiva do uso precoce (intra-operatório) e tardio (pós-operatório) do balão intra-aórtico. Foram estudados 130 pacientes do Texas Heart Institute, no período de janeiro a dezembro de 1987, sendo 103 do sexo masculino e 27 do sexo feminino; a idade média dos pacientes foi de 61,5 ± 10,76 (14 a 84) anos, a média de peso foi de 75,5 ± 16,6 (42 a 134) kg, a média de superfície corpórea de 1,87 ± 0,24 (1,08 a 2,60)m2 e um tempo médio de circulação extracorpórea de 98 (33 a 299) minutos e de pinçamento aórtico de 49 (10 a 122) minutos. O balão intra-aórtico foi usado em todos os pacientes por baixo débito cardíaco, em 4 associado a disrritmias retratarias a tratamento clínico e 2 por parada cardiorrespiratória, por períodos que variaram de 15 minutos a 256 horas. Os pacientes foram divididos em 6 grupos; Grupo I, pacientes que receberam o balão intra-aórtico no intra-operatório (precoce), Grupo II, pacientes que receberam o balão intra-aórtico no pós-operatório (tardio), Grupo III (pacientes com idade igual ou inferior a 65 anos), Grupo IV (pacientes com idade superior a 65 anos), Grupo V (pacientes com tempo de circulação extracorpórea de até 120 minutos) e Grupo VI (pacientes com tempo de circulação extracorpórea superior a 120 minutos) e 2 subdivisões nos grupos V e VI, subgrupo 1/. (pacientes com tempo de pinçamento de aorta menor ou igual a 60 minutos) e subgrupo 2/. (pacientes com tempo de pinçamento de aorta maior que 60 minutos). Dos 130 pacientes submetidos ao BIA, 81 (62.3%) sobreviveram e 49 (37,7%) pacientes faleceram; destes, 38 (36,2%) pertenciam ao Grupo I e 11 (44%) ao Grupo II. A sobrevida foi maior no Grupo III (68,6%) e menor no Grupo IV (51%) com p<0,05 mostrando uma diferença estatística com relação à mortalidade no grupo mais idoso. Observamos, também, que, a fração de ejeção, tempo de pinçamento de aorta, não apresentaram diferença estatística significante com relação à mortalidade. Porém a mortalidade com relação ao tempo de CEC foi altamente positiva (p<0,01) para tempo de CEC maior que 120 minutos. Em nossa casuística, o índice de complicações foi de 4,6% (6 pacientes). Os dados deste estudo sugerem que o balão intra-aórtico é efetivo como método de suporte circulatório, em pacientes com baixo débito cardíaco pós circulação extracorpórea, havendo tendência a resultados melhores quando usado precocemente; foi observada influência da idade e do tempo de CEC, sugerindo que métodos mais eficientes de assistência circulatória devam ser usados nos pacientes mais idosos e nos pacientes com síndrome de baixo débito cardíaco pós-cirurgia cardíaca com tempos de CEC maiores que 120 minutos.
Full text available only in portuguese PDF format.

REFERENCES

1. BACIEWICZ Jr., F. A.; KAPLAN, B. M.; MURPHY, T. E.; NEIMAN, H. L. - Bilateral renal artery thrombotic occlusion: a unique complication following removal of a transthoracic intra-aortic balloon. Ann. Thorac. Surg., 33: 631-634, 1982. [MedLine]

2. BARBERO-MARCIAL, M. - Circulação assistida. In: ZERBINI, E. J. (ed.) Clínica Cirúrgica Alípio Corrêa Netto. 3. ed. São Paulo, Sarvier, 1974. p. 436-444.

3. BARBERO-MARCIAL, M. - Síndrome de baixo débito cardíaco. In: STOLF, N. A. G. & ZERBINI, E. J. Pós-operatório em cirurgia cardíaca. São Paulo, Sarvier, 1979. p. 81-89.

4. BARBERO-MARCIAL, M.; KAWABE, L.; DALLAN, L. A. O.; RAMIRES, J. A. F.; BELLOTTI, G.; BITTENCOURT, D.; VERGINELLI, G.; PILEGGI, F.; DECOURT, L. V.; ZERBINI, E. J. - Contrapulsação com balão intra-aórtico em pacientes portadores de infarto agudo do miocardio. Rev. Hosp. Clin. Fac. Med. (São Paulo), 34: 171-176, 1979. [MedLine]

5. BREGMAN, D. A. - Clinical experience with percutaneous intra-aortic balloon pumping. Cardiovasc. Dis. Bull. Tex. Heart Inst, 7: 318-324, 1980.

6. BREGMAN, D. & KASKEL, P. - Advances in percutaneous intra-aortic ballon pumping. Critical Care Clin., 2: 221-236, 1986.

7. BREGMAN, D.; PARODI, E. N.; HAUBERT, S. M.; SZARNICKI, R.; EDIE, R. N.; SPOTNITZ, H. M.; BOWMAN, Jr. O. B.; REEMSTMA, K.; MALM, J. R. - Unidirectional intra-aortic balloon pumping and counterpulsation during open heart surgery. In: BREGMAN, D. A. (ed.) Mechanical support of the failing heart and lungs. New York, Appleton-Century-Crofts, 1977. p. 17-46.

8. BUCKLEY, M. J. - Intra-aortic balloon circulatory support. In: GRILLO, H. C.; AUSTEN, W. G.; WILKINS, Jr. E. W.; MATHISEN, D. J.; VLAHAKES, G. J. (eds.) Current Therapy in cardiothoracic surgery. Philadelphia, B. C. Decker, 1989, p. 318-319.

9. COSTA, A. R.; TEIXEIRA, G.; PEREIRA, E.; PRATES, P.; KALIL, R. A. K.; LUCHESE, F. A.; SANT'ANNA, J. R.; NESRALLA, I. A. - Emprego do balão intra-aórtico em cirurgia cardíaca. Arq. Bras. Cardiol., 59 (Supl. 2): 94, 1992. (Resumo) [MedLine]

10. DILELLO, F. - Results of intraaortic balloon pumping after cardiac sugery: experience with the Percor Balloon catheter. Ann. Thorac. Surg., 46: 442-447, 1988.

11. GOLDMAN, B. S.; HILL, T. J.; ROSENTHAL, G. A.; SCULLY, H. E.; WEISEL, R. D.; BAIRD, R. J. - Complications associated with use of the intra-aortic balloon pump. Canadian J. Surg., 25: 153-156, 1982.

12. KANTROWITZ, A.; TJONNELAND, S.; FREED, P. S. - Initial clinical experience with intra-aortic balloon pumping in cardogenic shock. J A M A, 203: 113-118, 1968.

13. KERN, M. J. - Intra-aortic balloon counterpulsation. Coron. Artery Dis., 2: 649-660, 1991.

14. KIRKLIN, J. W. & BARRAT-BOYES, B. G. - Cardiac Surgery. New York. John Wiley & Sons, 1986. p. 150-166.

15. KOHLER, I.; BOROWSKY, A.; KORB, H.; OZEL, O.; DE VIVE, E. R. - Suporte cardiocirculatorio (SC) em sindrome de baixo débito cardíaco (SBDC) em pósoperatório (PO) de cirurgia cardíaca. Arq. Bras. Cardiol., 61 (Supl. 2): 186, 1993. (Resumo) [MedLine]

16. MARKS, J. D.; KARWANDE, S. V.; RICHENBACHER, W. E.; JONES, K. W.; DOTY, D. B.; MILLAR, R. C; O'CONNELL, J. B.; RENLUND, D. G.; BRISTOW, M. R.; PENTALOS, G. M. - Perioperative mechanical circulatory support for transplantation. J. Heart Lung Transplant, 11: 117-128, 1992. [MedLine]

17. MARTIN III, R. S.; MONCURE, A. C.; BUCKLEY, M. J.; AUSTEN, W. G.; AKINS, C.; LEIBACK, R. C. - Complications of percutaneous intra-aortic balloon insertion. J. Thorac. Cardiovasc. Surg., 85: 186-190, 1983. [MedLine]

18. MATTOX, K. L. & BEALL, A. C. - Application of portable cardiopulmonary bypass to emergency instrumentation. Med. Instrum., 11: 347-349, 1979.

19. MILLER, R. R. - Combined dopamine and nitroprusside therapy in congestive heart failure. Circulation, 55: 881,1977. apud BRAUNWALD, E. - Heart Disease. 3. ed. Philadelphia, 1988. Cap. 17, p. 485-543: Management of Heart Failure.

20. MOULOPOULUS, S. D. - Mechanical assistence in cardiogenic shock. J. Biomater. Appl., 4: 544-551, 1989.

21. MURAD, H.; NASCIMENTO, F. J. V.; NENO, A. C. A.; ALBUQUERQUE, P. C. C.; QUEIROZ, A. D.; FILHO, G. R.; MURAD, M.; SANTOS, M. V. N.; OLIVEIRA, D. A. - Utilização do balão intra-aórtico no pré e pós-operatório de cirurgia cardiaca: experiencia do Hospital São Vicente de Paulo - RJ. Arq. Bras. Cardiol., 57: 57, 1991. (Resumo) [MedLine]

22. NAWA, S.; YAMADA, M.; TERAMOTO, S. - Evaluation of conventional circulatory assist devices: intraaortic balloon pumping, venoarterial bypass, and extracorporeal membrane oxigenation. Chest., 95: 261-266, 1989. [MedLine]

23. PENNINGTON, D. G.; SWARTZ, M.; CODD, J. E.; MERJAVY, J. P.; KAISER, G. C. - Intraaortic balloon pumping in cardiac surgical patients: a nine-year experience. Ann. Thorac. Surg., 36: 125-131, 1983. [MedLine]

24. PHILLIPS, S. J. - Percutaneous cardiopulmonary bypass and innovations in clinical counterpulsation. Critical Care Clinics, 2: 297-318, 1986. [MedLine]

25. SCOTT, M. L. & MILAN, J. R. - Long-term management of intra-aortic balloon pump dependent patient. Angiol. J. Vasc. Dis., March: 192-196, 1983.

26. SLOGOFF, S. - Dopamine, dobutamine, digitalis, and diuretics during intraaortic balloon support. Cardiovasc. Dis. Bull. Texas Heart Inst, 7: 325-330, 1980.

27. SMITH, T. W.; BRAUNWALD, E.; KELLY, R. A. - Management of heart failure. In: BRAUNWALD, E. (ed.) Heart Disease. 3. ed. Philadelphia, W. B. Saunders, 1988. p. 485.

28. SPENCER, F. C.; EISEMAN, B.; TRINKLE, J. K.; ROSSI, N. P. - Assisted circulation for cardiac failure following intracardiac surgery with cardiopulmonary bypass. J. Thorac. Cardiovasc. Surg., 49: 56-73, 1965. [MedLine]

29. SPOTNITZ, H. M.; BERMAN, M. A.; REIS, R. L.; EPSTEIN, S. E. - The effects of synchronized counterpulsation of the pulmonary artery on right ventricular hemodynamics. J. Thorac. Cardiovasc. Surg., 61: 167-170, 1971.

30. STURM, J. T.; FUHRMAN, T. M.; IGO, S. R.; HOLUB, D. A.; McGEE, M. G.; FUQUA, J. M.; NORMAN, J. C. - Quantitative indices of intra-aortic balloon (IABP): dependence during pós-infarction cardiogenic shock. Transac. Am. Soc. Artif. Intern. Organs., 4: 8-12, 1980.

31. TIEFENBRUNN, A. J. - Circulatory support devices in clinical cardiology. Coron. Artery Dis., 2: 637-639, 1991.

32. UGORJI, C. C.; TURNER, S. A.; McGEE, M. G.; FUHRMAN, T. M.; COOLEY, D. A.; NORMAN, J. C. - Transascending aortic intraaortic balloon insertion with delayed sternal closure: a retrospective analysis. Cardiovasc. Dis. Bull. Texas Heart Inst., 7: 307-315, 1980.

33. WALDHAUSEN, J. A. & PIERCE, W. S. - Johnson's Surgery of the chest. 5. ed. Chicago, 1985. p. 269-275.

Article receive on Thursday, August 31, 1995

CCBY All scientific articles published at www.bjcvs.org are licensed under a Creative Commons license

Indexes

All rights reserved 2017 / © 2024 Brazilian Society of Cardiovascular Surgery DEVELOPMENT BY