Jeffrey Shi Kai ChanI; Sukhdeep SinghII; Peter EriksenIII; Lok Him TsuiIV; Amer HarkyV; Cardiovascular Evidence Review Collaboration (CvERC)
DOI: 10.21470/1678-9741-2020-0146
ABSTRACT
Objectives: Bicuspid aortic valve (BAV) is an important aetiology of aortic stenosis and the use of transcatheter aortic valve implantation (TAVI) has not been fully explored in this cohort. This systematic review and meta-analysis compared the outcomes of TAVI in stenotic BAV against tricuspid aortic valve (TAV).AKI = Acute kidney injury
AS = Aortic stenosis
BAV = Bicuspid aortic valve
CI = Confidence interval
PPM = Permanent pacemaker
PRISMA = Preferred Reporting Items for Systematic Reviews and Meta-Analyses
RR = Risk ratio
SPSS = Statistical Package for the Social Sciences
TAVI = Transcatheter aortic valve implantation
TAV = Tricuspid aortic valve
TSA = Trial sequential analysis
WMD = Weighted mean differences
INTRODUCTION
Transcatheter aortic valve implantation (TAVI) is a well-established treatment strategy in patients with severe symptomatic aortic stenosis with high surgical risk for conventional aortic valve replacement[1,2]. However, this recommendation was based on clinical trials that excluded patients with bicuspid aortic valve (BAV)[1,2], a common cardiac anomaly present in 0.5-2% of the general population and associated with the development of aortic stenosis (AS) requiring intervention[3]. Generally, patients with BAV have larger annular dimensions, may have variable coronary anatomy, more calcified, bulky and irregular aortic valve leaflets, and altered aortic geometry and blood flow[4,5]. These differences can complicate the accurate device delivery and apposition of the prosthetic valve during TAVI[6-8]. However, outcomes of TAVI in patients with BAV using new-generation valves have shown promising results, with less paravalvular leak and better postprocedural outcomes than early-generation valves[9-11]. A significant number of centres around the world have also started performing TAVI on stenotic BAV patients. This systematic review and meta-analysis thus sought to thoroughly examine the literature to compare the outcomes of using TAVI in BAV replacement.
METHODS
This systematic review and meta-analysis was reported according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement, and was conducted according to The Cochrane Handbook for Systematic Reviews of Interventions[12,13]. Electronic searches were performed on PubMed, Scopus, MEDLINE, and EMBASE from their inception up till October 2019 to identify all publications that reported the use of TAVI in patients with BAV. ClinicalTrials.gov was also searched to identify ongoing or unpublished clinical trials. The search string used was "TAVI" OR "valve implantation" OR "percutaneous" AND "bicuspid valve" OR "bicuspid aorta" OR "bicuspid aortic valve" OR "aortic stenosis". Reference lists of identified papers were searched manually to identify other eligible studies.
Inclusion and Exclusion Criteria
Only studies written in English comparing TAVI in at least five patients with stenotic BAV and tricuspid aortic valve (TAV) were included. Non-comparative studies, studies with less than five patients, and studies including re-do valve-in-valve, tricuspid valve or aortic regurgitation were excluded. Articles were screened by three reviewers (JSKC, PE, LHT). All selected articles were systematically assessed with inclusion and exclusion criteria. Conflicts over inclusion were resolved by an independent reviewer (AH). All included studies were critically appraised using the Newcastle-Ottawa Scale.
Data Extraction and Reported Outcomes
Summary estimates were manually extracted by three reviewers (SS, PE, LHT). When there were duplicate data, only the most updated data were included. Conflicts over data extraction were resolved by an independent reviewer (JSKC). Primary endpoints included 30-day and 1-year mortality. Secondary endpoints included post-operative stroke, AKI, and need for permanent pacemaker (PPM) implantation. Other baseline, operative and post-operative characteristics were also extracted.
Statistical Analysis
Risk ratio (RR [95% confidence interval (CI)]) or weighted mean differences (WMD [95% CI]) were used as summary measures for primary endpoints. Random effects model was used with the Mantel-Haenszel test or inverse variance analysis, as appropriate. Heterogeneity was assessed by the chi-square test and the I2 statistic, for which values >0.40 were considered to imply significant heterogeneity. Sensitivity analysis was performed by removing studies individually from the analysis.
Trial sequential analysis (TSA) was performed on all outcomes using a combination of sample size and event size. O'Brien-Fleming α-spending function was used to adjust the Z-score threshold. Studies with 0 events were handled by adding a constant (1) to both the intervention and control arm. Required information size was estimated from all included studies reporting the analysed variables and incidences calculated from included patients, with a permissible two-sided type 1 error of 5% and type 2 error of 20%. TSA was performed using the Copenhagen trial unit TSA software version 0.9.5.10 beta.
All P-values were 2-sided, with P<0.05 considered significant. Statistical analyses were performed using Review Manager V.5.3 (Copenhagen: The Nordic Cochrane Centre, The Cochrane Collaboration, 2014) and SPSS software version 25.0 (IBM Corp, Armonk, New York, USA).
RESULTS
Eight studies were deemed eligible for inclusion in this meta-analysis (Figure 1)[14-21]. The studies were excluded due to the lack of reports on TAV and BAV cohorts in the same article, in case series with less than 5 patients or in single cohort studies. The articles represented a total of 3996 patients (917 with BAV and 3079 with TAV, Table 1). An electronic search on ClinicalTrials.gov also identified one relevant single-blinded randomized controlled trial. The study is expected to be completed in 2023 (ClinicalTrials.gov identifier: NCT02541877). Results of the critical appraisal by the Newcastle-Ottawa Scale were summarized in Table 2. Each asterisk represents 1 point, with ≥7 out of 9 points unlikely to have a significant risk of bias. The assessment results showed that all studies were unlikely to have a significant risk of bias.
Article | Year of publication | Study type | BAV (n=1077) |
TAV (n=4165) |
Summary of article |
---|---|---|---|---|---|
Aalei-Andabili et al.[14] | 2018 | Retrospective cohort | 32 | 96 | TAVI in patients with BAV provides both comparable immediate and mid-term outcomes with TAV and is feasible. |
Bauer et al.[15] | 2014 | Prospective cohort | 38 | 1357 | In selected patients, TAVI for BAV can provide satisfactory results. Risk of relevant AR appears greater in patients with BAV; however, both 30-day and 1-year mortalities were not elevated in relation to TAV. |
De Biase et al.[16] | 2018 | Prospective cohort | 83 | 166 | More complex anatomy associated with BAV at baseline leads to lower device success rates, but this is not associated with higher 30-day mortalities. |
Kochman et al.[17] | 2014 | Prospective cohort | 28 | 84 | Selected high-risk BAV patients can be successfully treated with TAVI and have similar outcomes to non-BAV patients. |
Liao et al.[18] | 2018 | Prospective cohort | 87 | 70 | TAVI for BAV looks safe and effective, with comparable bioprosthetic valve functionality compared to TAV. |
Liu et al.[19] | 2015 | Prospective cohort | 15 | 25 | There was no difference between the success of device, 30-day mortality or the 30-day combined endpoints between TAVI in BAV and TAV. |
Sannino et al.[20] | 2017 | Retrospective cohort | 88 | 735 | TAVI appears safe and effective in BAV, with no differences in post-procedure mortality, or 30-day cardiovascular mortality compared to patients with TAV. |
Yoon et al.[21] | 2017 | Prospective and retrospective cohort | 546 | 546 | TAVI in BAV was associated with similar prognosis but had lower device success. Procedural differences occurred with early devices, but not with new generation ones. |
Author | Selection | Comparability | Outcome | |||||||
---|---|---|---|---|---|---|---|---|---|---|
Representation of patients with bicuspid aortic valve | Selection of patients with tricuspid aortic valve | Ascertainment of exposure | Demonstration that the outcome of interest was not present at the start of the study | Study controls for patient age =✹ | Study controls for preoperative cardiac function and cardiovascular co-morbidities =✹ | Assessment of outcomes | Follow-up long enough for outcomes to occur | Adequacy of follow-up of cohorts | Overall quality score (maximum=9) | |
Aalaei-Andabili et al.[14] | ✹ | ✹ | ✹ | ✹ | ✹✹ | ✹ | ✹ | 8 | ||
Bauer et al.[15] | ✹ | ✹ | ✹ | ✹ | ✹✹ | ✹ | ✹ | 8 | ||
De Biase et al.[16] | ✹ | ✹ | ✹ | ✹ | ✹✹ | ✹ | ✹ | 8 | ||
Kochman et al.[17] | ✹ | ✹ | ✹ | ✹ | ✹✹ | ✹ | ✹ | ✹ | 9 | |
Liao et al.[18] | ✹ | ✹ | ✹ | ✹ | ✹✹ | ✹ | ✹ | ✹ | 9 | |
Liu et al.[19] | ✹ | ✹ | ✹ | ✹ | ✹✹ | ✹ | ✹ | 8 | ||
Sannino et al.[20] | ✹ | ✹ | ✹ | ✹ | ✹✹ | ✹ | ✹ | ✹ | 9 | |
Yoon et al.[21] | ✹ | ✹ | ✹ | ✹ | ✹✹ | ✹ | ✹ | ✹ | 9 |
The baseline characteristics were summarized in Table 3. The BAV cohort was significantly younger (WMD -0.89 years [-1.60 years, -0.17 year], P=0.02), and had a higher trans-aortic valve gradient (WMD 1.73 mmHg [0.31 mmHg, 3.16 mmHg], P=0.02), and a larger ascending aortic diameter (WMD 3.92 mm [3.02 mm, 4.83 mm], P<0.0001). All other baseline characteristics were not significantly different.
Baseline variable | BAV (n=917) | TAV (n=3079) | OR/WMD [95% CI] | P-value |
---|---|---|---|---|
Age (years±SD) | 76.8±7.43 | 78.5±7.12 | WMD -0.89 [-1.60, -0.17] | 0.02* |
Male (%) | 562/917 (61.3) | 1554/3079 (50.5) | OR 1.10 [0.93, 1.31] | 0.27 |
Diabetes mellitus (%) | 226/917 (24.6) | 980/3079 (31.8) | OR 1.00 [0.82, 1.21] | 0.98 |
Previous stroke (%) | 130/917 (14.2) | 357/3079 (11.6) | OR 1.20 [0.94, 1.54] | 0.15 |
Peripheral arterial disease (%) | 177/834 (21.2) | 690/2913 (23.7) | OR 0.95 [0.67, 1.34] | 0.76 |
Hypertension (%) | 603/879 (68.6) | 1293/1722 (75.1) | OR 0.95 [0.78, 1.16] | 0.61 |
COPD (%) | 194/834 (23.3) | 660/2913 (22.7) | OR 1.05 [0.84, 1.32] | 0.67 |
Ischaemic heart disease (%) | 175/339 (51.6) | 1472/2437 (60.4) | OR 0.99 [0.76, 1.29] | 0.92 |
Previous CABG (%) | 75/710 (10.6) | 340/2178 (15.6) | OR 0.85 [0.62, 1.17] | 0.33 |
Previous PCI (%) | 180/797 (22.6) | 709/2248 (31.5) | OR 0.89 [0.71, 1.11] | 0.31 |
NYHA class III-IV (%) | 632/797 (79.3) | 1874/2248 (83.4) | OR 1.09 [0.86, 1.37] | 0.48 |
Atrial fibrillation (%) | 49/273 (17.9) | 189/996 (19.0) | OR 0.94 [0.64, 1.38] | 0.74 |
EuroSCORE (%±SD) | 17.4±10.5 | 19.4±12.6 | WMD -1.15 [-2.63, 0.33] | 0.13 |
STS score (%±SD) | 6.10±3.88 | 6.53±3.98 | WMD 0.06 [-0.28, 0.40] | 0.73 |
Aortic valve area (cm2±SD) | 0.629±0.173 | 0.660±0.239 | WMD -0.03 [-0.06, 0.01] | 0.11 |
Trans-aortic valve gradient (mmHg±SD) | 52.6±17.3 | 49.8±16.1 | WMD 1.73 [0.31, 3.16] | 0.02* |
Aortic annular diameter (mm±SD) | 23.6±4.42 | 24.3±3.72 | WMD -0.23 [-1.63, 1.17] | 0.74 |
Left ventricular ejection fraction (%±SD) | 51.3±14.2 | 53.6±13.0 | WMD -2.60 [-5.57, 0.37] | 0.09 |
Ascending aortic diameter (mm±SD) | 38.4±5.00 | 34.2±3.63 | WMD 3.92 [3.02, 4.83] | <0.0001* |
Operative outcomes were summarized in Table 4 and postoperative outcomes in Table 5. All primary and secondary outcomes were not significantly different and did not have significantly heterogeneous data. These included postoperative stroke (RR 1.22 [0.69, 2.14], P=0.49; I2=0, chi-square=3.29, P=0.86; Figure 2), AKI (RR 1.78 [0.83, 3.85], P=0.14; I2=0, chi-square=0.39, P=0.82; Figure 3), PPM requirement (RR 0.98 [0.82, 1.18], P=0.86; I2=0, chi-square=6.96, P=0.43; Figure 4), 30-day mortality (RR 1.17 [0.78, 1.73], P=0.44; I2=0, chi-square=2.11, P=0.95; Figure 5), and 1-year mortality (RR 0.89 [0.68, 1.17], P=0.41; I2=0, chi-square=3.90, P=0.42; Figure 6). However, the BAV cohort had considerable higher rates of significant aortic regurgitation (more than grade 2) postoperatively (RR 1.53 [1.17, 1.99], P=0.002). All other operative and postoperative outcomes were not significantly different. Sensitivity analysis revealed that the individual removal of studies from the analysis would not affect the statistical significance of the meta-analytical results of any variable.
BAV (n=917) | TAV (n=3079) | OR [95% CI] | P-value | ||
---|---|---|---|---|---|
Vascular access route | Transfemoral (%) | 689/829 (83.1) | 2015/2330 (86.5) | 0.82 [0.52, 1.28] | 0.37 |
Transapical (%) | 22/186 (11.8) | 216/2272 (9.51) | 1.11 [0.68, 1.80] | 0.68 | |
Transaortic (%) | 6/284 (2.11) | 48/2463 (1.95) | 1.14 [0.42, 3.08] | 0.80 | |
Others (%) | 4/201 (1.99) | 57/2297 (2.48) | 0.91 [0.33, 2.47] | 0.85 | |
Type of valve used | Sapien XT (%)x | 175/606 (28.9) | 218/726 (30.0) | 1.01 [0.79, 1.29] | 0.95 |
CoreValve (%) | 256/746 (34.3) | 1391/2178 (63.9) | 1.06 [0.62, 1.83] | 0.82 | |
Venus A (%) | 64/102 (62.8) | 60/95 (63.2) | 0.81 [0.36, 1.83] | 0.61 | |
SAPIEN 3 (%) | 232/699 (33.2) | 507/2165 (23.4) | 1.38 [0.76, 2.53] | 0.29 | |
Lotus (%) | 46/629 (7.31) | 51/712 (7.16) | 0.94 [0.62, 1.43] | 0.78 | |
EvolutR (%) | 39/629 (6.20) | 114/712 (16.4) | 0.49 [0.06, 4.13] | 0.51 |
BAV (n=917) | TAV (n=3079) | RR/WMD [95% CI] | P-value | |
---|---|---|---|---|
Device success (%) | 712/830 (85.8) | 2812/3009 (93.5) | RR 0.96 [0.90, 1.03] | 0.24 |
Bleeding (%) | 83/764 (10.9) | 183/1460 (12.5) | RR 1.00 [0.72, 1.39] | 0.98 |
Conversion to open surgery (%) | 12/746 (1.61) | 16/2178 (0.73) | RR 2.89 [0.53, 15.83] | 0.22 |
Vascular complications (%) | 38/802 (4.74) | 105/2817 (3.73) | RR 1.02 [0.66, 1.56] | 0.94 |
Trans-aortic valve gradient (mmHg±SD) | 9.74±5.96 | 9.78±5.15 | WMD 0.09 [-0.36, 0.54] | 0.70 |
AR more than grade 2 (%) | 88/815 (10.8) | 303/2984 (10.2) | RR 1.53 [1.17, 1.99] | 0.002* |
Stroke (%) | 23/917 (2.51) | 88/3179 (2.77) | RR 1.22 [0.69, 2.14] | 0.49 |
Acute kidney injury (%) | 14/677 (2.07) | 23/1390 (1.66) | RR 1.78 [0.83, 3.85] | 0.14 |
PPM requirement (%) | 158/917 (17.2) | 776/3079 (25.2) | RR 0.98 [0.82, 1.18] | 0.86 |
30-day mortality (%) | 43/917 (4.69) | 210/3079 (6.82) | RR 1.17 [0.78, 1.73] | 0.45 |
1-year mortality (%) | 76/787 (9.66) | 414/2792 (14.8) | RR 0.89 [0.68, 1.17] | 0.41 |
TSA was performed for 30-day mortality, 1-year mortality, PPM requirement, and postoperative stroke. For 30-day mortality (Figure 7), 1-year mortality (Figure 8) and PPM requirement (Figure 9), the cumulative Z curve of the study data crossed the futility boundary and/or the trial sequential monitoring boundary, indicating that the meta-analysis of these variables was conclusive. However, the cumulative Z curves for postoperative stroke (Figure 10) did not cross any of the boundaries, indicating that meta-analyses of these variables were inconclusive, and further studies were required. TSA was not possible for AKI due to the small sample and event sizes.
DISCUSSION
TAVI has not been fully explored in patients with BAV, and the latest guidelines did not fully support the use of TAVI in BAV patients[22,23]. We thus evaluated the existing evidence for the safety of TAVI in stenotic BAV compared to that in TAV. In this study, a total of eight articles with 917 stenotic BAV and 3079 stenotic TAV patients were meta-analysed. We showed no significant difference in primary and secondary outcomes, including AKI, PPM requirement, stroke, and 30-day mortality. TSA results confirmed that 30-day and 1-year mortality, as well as PPM requirements, were not significantly different between BAV and TAV, while more evidence was required for stroke. As such, TAVI is largely safe for clinical use in stenotic BAV patients.
However, BAV was associated with a considerable higher rate of significant aortic regurgitation postoperatively (P=0.002). It should be noted that if studies that started recruitment before 2012 were excluded, i.e. Bauer et al.[15], Kochman et al.[17] and Yoon et al.[21], the difference in the rates of significant postoperative aortic regurgitation would become statistically insignificant (RR 1.31 [0.67, 2.54], P=0.43; I2=0, chi-square=0.33, P=0.85). This reflected that there was a learning curve for TAVI and the inclusion of older, earlier procedures skewed the data-in fact, Yoon et al.[21] included procedures performed as early as 2005. A plot between the year of publication and the rate of significant postoperative aortic regurgitation showed a decreasing trend across the years (Figure 11). The idea of a learning curve for TAVI has been suggested by others and, therefore, having contemporary data is important[24,25].
Previous meta-analyses on the same topic were published by Takagi et al.[26] and Ueshima et al.[27]. Our results largely agree with these studies. Nonetheless, to the best of the authors' knowledge, this was the first meta-analysis to utilize TSA in evaluating TAVI in BAV versus TAV[28]. TSA builds on the simple fact that the conclusiveness of any evidence increases with the sample size studied, and that the stronger a true effect is, the fewer the number of subjects that need to be studied for the observed effects to be conclusive. TSA results are presented as plots: whenever the Z-score curve (blue line) crosses either the statistical significance boundary (outer oblique boundaries), the futility boundary (inner oblique boundaries), or the trial sequential monitoring boundary (vertical boundary), the results may be considered stable and conclusive. Our results justified further studies designed to evaluate postoperative stroke with adequate power. In fact, with almost 4000 patients included in comparative studies, TSA should be encouraged in subsequent meta-analyses to check if a definitive conclusion has been reached.
This meta-analysis had several limitations. First, it was possible that the manual electronic search had lost articles eligible for inclusion in this study, as well as studies written in non-English languages. Second, late mortality (>1-year follow-up) could not be analysed since only 1 study (Yoon et al.[21] followed patients for more than 1 year with adequate data for analysis. These might mean that long-term data of TAVI in TAV were likely not extrapolatable to BAV patients. This is an important consideration for future studies.
Third, subgroup analysis by the BAV subtype was not possible, as none of the included studies reported separate data by BAV subtypes. This has been considered a key parameter that should be assessed in preoperative imaging of BAV patients considered for TAVI. Similarly, subgroup analysis by the type of valve used was not possible due to studies including a mixture of different valves, often from different generations. Reporting outcomes by the type of valve could give a better understanding of whether the benefits of newer generation valves in TAV were applicable in BAV. These should be considered in future studies.
REFERENCES
1. Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, et al. Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med. 2010;363(17):1597-607. doi:10.1056/NEJMoa1008232.
2. Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW, Svensson LG, et al. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med. 2011;364(23):2187-98. doi:10.1056/NEJMoa1103510.
3. Michelena HI, Desjardins VA, Avierinos JF, Russo A, Nkomo VT, Sundt TM, et al. Natural history of asymptomatic patients with normally functioning or minimally dysfunctional bicuspid aortic valve in the community. Circulation. 2008;117(21):2776-84. doi:10.1161/CIRCULATIONAHA.107.740878.
4. Watanabe Y, Chevalier B, Hayashida K, Leong T, Bouvier E, Arai T, et al. Comparison of multislice computed tomography findings between bicuspid and tricuspid aortic valves before and after transcatheter aortic valve implantation. Catheter Cardiovasc Interv. 2015;86(2):323-30. doi:10.1002/ccd.25830.
5. Mahadevia R, Barker AJ, Schnell S, Entezari P, Kansal P, Fedak PW, et al. Bicuspid aortic cusp fusion morphology alters aortic three-dimensional outflow patterns, wall shear stress, and expression of aortopathy. Circulation. 2014;129(6):673-82. doi:10.1161/CIRCULATIONAHA.113.003026.
6. Zegdi R, Lecuyer L, Achouh P, Didier B, Lafont A, Latrémouille C, et al. Increased radial force improves stent deployment in tricuspid but not in bicuspid stenotic native aortic valves. Ann Thorac Surg. 2010;89(3):768-72. doi:10.1016/j.athoracsur.2009.12.022.
7. Zhao ZG, Jilaihawi H, Feng Y, Chen M. Transcatheter aortic valve implantation in bicuspid anatomy. Nat Rev Cardiol. 2015;12(2):123-8. doi:10.1038/nrcardio.2014.161.
8. Himbert D, Pontnau F, Messika-Zeitoun D, Descoutures F, Détaint D, Cueff C, et al. Feasibility and outcomes of transcatheter aortic valve implantation in high-risk patients with stenotic bicuspid aortic valves. Am J Cardiol. 2012;110(6):877-83. doi:10.1016/j.amjcard.2012.04.064.
9. Yoon SH, Lefèvre T, Ahn JM, Perlman GY, Dvir D, Latib A, et al. Transcatheter aortic valve replacement with early- and new-generation devices in bicuspid aortic valve stenosis. J Am Coll Cardiol. 2016;68(11):1195-205. doi:10.1016/j.jacc.2016.06.041. [MedLine]
10. Husser O, Pellegrini C, Kessler T, Burgdorf C, Thaller H, Mayr NP, et al. Outcomes after transcatheter aortic valve replacement using a novel balloon-expandable transcatheter heart valve: a single-center experience. JACC Cardiovasc Interv. 2015;8(14):1809-16. doi:10.1016/j.jcin.2015.08.014.
11. Yoon SH, Sharma R, Chakravarty T, Kawamori H, Maeno Y, Miyasaka M, et al. Clinical outcomes and prognostic factors of transcatheter aortic valve implantation in bicuspid aortic valve patients. Ann Cardiothorac Surg. 2017;6(5):463-72. doi:10.21037/acs.2017.09.03. [MedLine]
12. Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gøtzsche PC, Ioannidis JP, et al. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate healthcare interventions: explanation and elaboration. BMJ. 2009;339:b2700. doi:10.1136/bmj.b2700.
13. Higgins JPT, Thomas J, Chandler J, Cumpston M, Li T, Page MJ, Welch VA (editors). Cochrane Handbook for Systematic Reviews of Interventions version 6.0 (updated July 2019). Cochrane, 2019. Available from
14. Aalaei-Andabili SH, Beaver TM, Petersen JW, Anderson RD, Karimi A, Thoburn E, et al. Early and midterm outcomes of transcatheter aortic valve replacement in patients with bicuspid aortic valves. J Card Surg. 2018;33(9):489-96. doi:10.1111/jocs.13775. [MedLine]
15. Bauer T, Linke A, Sievert H, Kahlert P, Hambrecht R, Nickenig G, et al. Comparison of the effectiveness of transcatheter aortic valve implantation in patients with stenotic bicuspid versus tricuspid aortic valves (from the German TAVI registry). Am J Cardiol. 2014;113(3):518-21. doi:10.1016/j.amjcard.2013.10.023.
16. De Biase C, Mastrokostopoulos A, Philippart R, Desroche LM, Blanco S, Rehal K, et al. Aortic valve anatomy and outcomes after transcatheter aortic valve implantation in bicuspid aortic valves. Int J Cardiol. 2018;266:56-60. doi:10.1016/j.ijcard.2018.01.018. [MedLine]
17. Kochman J, Huczek Z, Scislo P, Dabrowski M, Chmielak Z, Szymanski P, et al. Comparison of one- and 12-month outcomes of transcatheter aortic valve replacement in patients with severely stenotic bicuspid versus tricuspid aortic valves (results from a multicenter registry). Am J Cardiol. 2014;114(5):757-62. doi:10.1016/j.amjcard.2014.05.063.
18. Liao YB, Li YJ, Xiong TY, Ou YW, Lv WY, He JL, et al. Comparison of procedural, clinical and valve performance results of transcatheter aortic valve replacement in patients with bicuspid versus tricuspid aortic stenosis. Int J Cardiol. 2018;254:69-74. doi:10.1016/j.ijcard.2017.12.013. [MedLine]
19. Liu XB, Jiang JB, Zhou QJ, Pu ZX, He W, Dong AQ, et al. Evaluation of the safety and efficacy of transcatheter aortic valve implantation in patients with a severe stenotic bicuspid aortic valve in a Chinese population. J Zhejiang Univ Sci B. 2015;16(3):208-14. doi:10.1631/jzus.B1500017.
20. Sannino A, Cedars A, Stoler RC, Szerlip M, Mack MJ, Grayburn PA. Comparison of efficacy and safety of transcatheter aortic valve implantation in patients with bicuspid versus tricuspid aortic valves. Am J Cardiol. 2017;120(9):1601-6. doi:10.1016/j.amjcard.2017.07.053. [MedLine]
21. Yoon SH, Bleiziffer S, De Backer O, Delgado V, Arai T, Ziegelmueller J, et al. Outcomes in transcatheter aortic valve replacement for bicuspid versus tricuspid aortic valve stenosis. J Am Coll Cardiol. 2017;69(21):2579-89. doi:10.1016/j.jacc.2017.03.017. [MedLine]
22. Baumgartner H, Falk V, Bax JJ, De Bonis M, Hamm C, Holm PJ, et al. 2017 ESC/EACTS guidelines for the management of valvular heart disease. Eur Heart J. 2017;38(36):2739-91. doi:10.1093/eurheartj/ehx391. [MedLine]
23. Nishimura RA, Otto CM, Bonow RO, Carabello BA, Erwin JP 3rd, Fleisher LA, et al. 2017 AHA/ACC focused update of the 2014 AHA/ACC guideline for the management of patients with valvular heart disease: a report of the American college of cardiology/American heart association task force on clinical practice guidelines. Circulation. 2017;135(25):e1159-95. doi:10.1161/CIR.0000000000000503. [MedLine]
24. Kaier K, Reinecke H, Schmoor C, Frankenstein L, Vach W, Hehn P, et al. Learning curves among all patients undergoing transcatheter aortic valve implantation in Germany: a retrospective observational study. Int J Cardiol. 2017;235:17-21. doi:10.1016/j.ijcard.2017.02.138 [MedLine]
25. Alli OO, Booker JD, Lennon RJ, Greason KL, Rihal CS, Holmes DR Jr. Transcatheter aortic valve implantation: assessing the learning curve. JACC Cardiovasc Interv. 2012;5(1):72-9. doi:10.1016/j.jcin.2011.09.014.
26. Takagi H, Hari Y, Kawai N, Kuno T, Ando T; ALICE (All-Literature Investigation of Cardiovascular Evidence) Group. Meta-analysis of transcatheter aortic valve implantation for bicuspid versus tricuspid aortic valves. J Cardiol. 2019;74(1):40-8. doi:10.1016/j.jjcc.2019.03.018. [MedLine]
27. Ueshima D, Nai Fovino L, Brener SJ, Fabris T, Scotti A, Barioli A, et al. Transcatheter aortic valve replacement for bicuspid aortic valve stenosis with first- and new-generation bioprostheses: a systematic review and meta-analysis. Int J Cardiol. 2020;298:76-82. doi:10.1016/j.ijcard.2019.09.003. [MedLine]
28. Yoon SH, Maeno Y, Kawamori H, Miyasaka M, Nomura T, Ochiai T, et al. Diagnosis and outcomes of transcatheter aortic valve implantation in bicuspid aortic valve stenosis. Interv Cardiol. 2018;13(2):62-5. doi:10.15420/icr.2018:8:2. [MedLine]
Authors' roles & responsibilities
JSKC Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; final approval of the version to be published
SS Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; final approval of the version to be published
PE Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; final approval of the version to be published
LHT Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; final approval of the version to be published
AH Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; final approval of the version to be published
Article receive on Wednesday, April 1, 2020
Article accepted on Friday, April 24, 2020